Standard Sol, Reagents & Buffers for PT6

All Myron L handheld instruments are factory-calibrated using NIST-traceable Standard Solutions with defined conductivity and ppm values. These Standard Solutions are produced under tightly controlled conditions, prepared with reagent-grade salts dissolved in deionized water of at least 5 megohm-cm purity. Each solution delivers ±1% accuracy, consistent with values published in the International Critical Tables and fully traceable to the National Institute of Standards and Technology (NIST).

The conductivity Standard Solutions and pH Buffers are used for factory calibration. Regular use of these solutions is recommended to ensure specified instrument accuracy. Frequency of conductivity recalibration depends upon use, but once every month should be sufficient for an instrument used daily. pH models, depending upon use, should be recalibrated with pH 7 Buffer every 1-2 weeks, and checked with pH 4 and/or 10 Buffers at similar intervals. pH Sensor Storage Solution is recommended for keeping the pH sensor wet. Myron L solutions are available in quart/1 ltr., gallon/3,8 ltr. and 2 oz./59 ml plastic bottles, ready to use. (Custom value standards only offered in gallon/3,8 ltr. size bottles.)

Sidebar Sidebar Sidebar

Conductivity instruments are a convenient way to determine the parts per million of total dissolved solids (ppm/TDS) in boilers, cooling towers, reverse osmosis systems, etc. Although the International Unit (Sl) of measuring conductivity is the microsiemens/ cm (also known as micromhos/cm), a direct reading in ppm/TDS is sometimes preferred.

Myron L conductivity instruments and monitor/controllers are calibrated to read in ppm/442, ppm/ NaCI, or microsiemens. All three values are listed on our Standard Solutions.