![Pyxis IK-765SS-O3 Ozone + pH Panel Solution Pyxis IK-765SS-O3 Ozone + pH Panel Solution](https://cdn11.bigcommerce.com/s-xxofluljba/images/stencil/448x448/products/66326/62078/IK-765SS-O3-Ozone-Panel-Solution__35321__02992.1707731520.jpg?c=1)
![Pyxis IK-765SS-O3 Ozone + pH Panel Solution Pyxis IK-765SS-O3 Ozone + pH Panel Solution](https://cdn11.bigcommerce.com/s-xxofluljba/images/stencil/160w/products/66326/62079/IK-765SS-O3-Ozone-pH-Panel__15933__92163.1707731520.jpg?c=1)
Pyxis
Checkout securely with:
The IK-765SS-SO3 is a dual-parameter inline water analyzer specifically designed as a ‘Turn-Key’ monitoring solution for clean water applications to monitor Sulfite in water commonly used in water and process applications for scavenging of Chlorine and Oxygen. The IK-765SS-SO3 offers highly accurate, real-time measurement, display and data-logging of Sulfite (SO3), pH and Temperature utilizing proprietary Pyxis Lab smart sensor technology, coupled with a Pyxis UC-80 touch screen display and data logging terminal. The IK-765 series is offered in a convenient and easy to integrate panel mounted format for rapid installation and simple maintenance.
The IK-765 Sulfite/pH analyzer is equipped with the propriety Pyxis ST-765SS-SO3 smart sensor design which is membrane-free and based on unique principles and incorporates Pyxis’ advanced technology in the field of bare-gold electrochemical detection. The sensor measures the Sulfite level and pH simultaneously while performing temperature and pH compensation for the measurement of the reducing agent based on conditions present in the application of use. The IK- 765SS-SO3 panel is also provided with the ST-007 stainless steel flow reservoir with inlet rotameter, digital flow meter and UC-80 display/data logging terminal prewired to the sensor in RS-485 Modbus format with fully integrated sensor diagnostics and calibration interface.
This unique platform with internal sensor compensation is factory calibrated by Pyxis Lab using precise Sulfite concentration standards in an Argon purged environment resulting in highly accurate Sulfite detection consistent with Titration, Colorimetry or Fluorescence methodology as high as 100ppm.
*NOTE* For boiler feedwater and blowdown sulfite measurement, the IK-765SS-SO3 panel must be installed downstream of a sample cooler.
All Pyxis IK-765 series analyzers require a small installation footprint, offer simple operation and maintenance and are specifically designed for use in chlorine or oxygen scavenger residual monitoring applications. The analyzer may also be integrated with the optional Pyxis PowerCloud™ 4G remote gateway, which can upload on-site analyzer data to a cloud server in real time and allow for additional sensor or device inputs for cloud access. Contact Pyxis Lab for details on PowerCloud™ and Pyxis Cloud™.
Item | IK-765SS-B Series |
Part Number | 42163 |
Sensor Body Material | 304SS |
Sulfite Range | 0.00–100.00ppm Sulfite |
pH Range | 0–14 |
Sulfite Precision | ±1.0mg/L or 1% of the Value with pH compensation up to 9.0+ |
pH Precision | ±0.01 |
Sample Operating Temperature | 4-40 °C (40-104 °F) |
Sample Inlet Pressure | 7.25-30psi (0.05-0.2MPa) |
Sensor Max Pressure | 100psi (6.9Bar) |
Sensor Response Time | T95≤60s – Oxidizer / T95≤5s pH |
Measurement Interval | Continuous |
Installation | ST-007 (316L) Flow Cell with PRV-Rotameter – Hall Effect Flow Meter Included |
ST-007 Min Flow Rate | 200mL/min |
ST-007 Max Flow Rate | 400mL/min |
ST-007 Sample Inlet | 1/4inch OD |
ST-007 Sample Outlet | 1/4 inch OD |
Panel Power Supply | 96–260VAC / 50-60Hz; 200W |
Panel Storage Temperature | -4–158 ºF (-20–70 ºC) |
Panel Operational Temperature | 32–122 ºF (0–50 ºC) |
UC-80 Display | 4.3inch LCD Color 480×272 Pixel Resolution / Resistive Touch |
UC-80 Input | RS-485 Modbus – RTU |
UC-80 Output | (2) 4-20mA / RS-485 Modbus RTU / Modbus-TCP |
UC-80 Data Storage | Built-In 1GB of RAM for Storing up to 1-Million Data/Event Records |
UC-80 USB | (1) USB Host for Data Downloading and Screen Upgrade |
UC-80 Relative Humidity | 20–90% (No Condensation) |
UC-80 Altitude | <6,561ft (<2,000m) |
Dimension (HxWxD) | 440H x 350W x 153D mm |
Approx. Weight | 15kg |
Wet Material | 316L Stainless Steel / Polycarbonate |
Rating | IP-65 Panel Display / IP-65 Sensor |
Selectivity | Non-Selective / Cross Sensitive to other Oxidizing Species |
Compliance | EPA 334.0 / ISO-7393 |
Regulation | CE Marked / RoHS |
Typical Electrode Service Life | 2 Years |
Electrode Warranty | 6 Months |
Sensor Body Warranty | 13 Months |
Downloads