Pyxis
Pyxis
HM-520SS Oil In Water Sensor (0–100ppm)
Checkout securely with:
The Pyxis HM-500 series inline fluorometer sensors measure the concentration of oil in water read as ppm Marine Crude-Offshore Oil. The sensors utilize an LED sourced UV-fluorescence methodology at 365nm wavelength and 410/470nm excitation. Pyxis HM-500 series sensors are uniquely designed with extra photoelectric components that also monitor the color and turbidity of the sample water. This proprietary feature enables the HM-500 series to automatically compensate for color and turbidity in the water sample eliminating interferences commonly associated with real-world waters, up to 150NTU. The Pyxis HM-500 series inline sensors have a short fluidic channel and can be easily cleaned. The fluidic and optical arrangement of the HM-500 series are designed to overcome shortcomings associated with other fluorometers that have a distal sensor surface or a long, narrow fluidic cell. Traditional inline fluorometers are susceptible to color/turbidity interference and fouling and can also be very difficult to clean. These unique features provide HM-500 series with a level of accuracy far greater than conventional inline oil in water sensors and enable users to conduct wireless inline sensor cleanliness diagnostics as a predictive measure via the uPyxis APP for mobile and desktop devices.
Features
- HM-500 Series provided with ST-001 inline tee assembly ¾” FNPT/CPVC
- HM-500SS Series offers 304-Stainless Steel ¾” FNPT for Harsh Applications
- Submersion Adapter Kit (MA-102S) Available for CPVC submersed applications
- Unique design with extra photo-electric components
- Automatically compensates for color and turbidity contamination
- Flexible/Advanced Communication 4-20mA Output and RS-485 Output
- Connectible to any conventional microprocessor controller, PLC or DCS
- Bluetooth Enabled when used with MA-WB or POWERPack Series Adapters
- Bluetooth Wireless Interface via uPyxis APP on all Mobile & Desktop devices
- uPyxis APP allows for sensor cleanliness diagnosis, calibration & configuration
- uPyxis APP allows users to write calibration to sensor for variety of oils as ppm
- Lower cost and higher accuracy vs conventional UV fluorometers for oil in water
Typical Applications
- Heat exchanger leakage detection
- Oil tank leakage detection
- Intake water monitoring (UF & RO Feedwater)
- Reinjection water monitoring
- Industrial wastewater monitoring
- Marine discharge monitoring
- Plant runoff / storm water monitoring
Specifications
Item | HM-520SS |
Part Number | 52126 |
Range | 0.5–100ppm |
Accuracy | 0.1ppm or ≤3% of Reading |
Method | UV-Fluorescence |
Emission Wavelength | 365nm |
Excitation Wavelength | 410nm/470nm |
Outputs | 4-20mA Analog Output, RS-485 Digital Output with Modbus Protocol |
Installation | 3/4″ FNPT Threaded Ports |
Cable Length | 5ft with IP67 Connectors |
Power Supply | 22–26VDC, 65mA |
Dimension | See Figure 4 in User Manual |
Weight | 2.5lbs (1130g) |
Material | 304 Stainless Steel |
Op. Temperature | 40–104 ºF (4–40 ºC) |
Storage Temperature | 20–140 ºF (-7–60 ºC) |
Pressure | Up to 290lbs |
Enclosure Rating | IP67 |
Regulation | CE |
Downloads