The SP-910 is a multi-parameter and multi-wavelength fluorometer, colorimeter and turbidimeter specifically designed for Municipal, Environmental and Industrial water analysis. For industrial water treatment applications utilizing fluorescent tracer chemistries, the SP-910 is ideally suited with integrated direct read testing for Fluorescein and PTSA. The SP-910 uses industry standard colorimetric reagents, as well as Pyxis Lab® reagents and provides colorimetric measurements at 7 LED wavelengths. The SP-910 is pre-calibrated for colorimetric measurements of analyses common in industrial water treatment and other water testing in the laboratory.
In multiple side-by-side validation and comparison studies, the SP-910 has proven to be statistically more accurate than other devices on the market. The SP-910 contains multiple methods complaint with USEPA guidelines for drinking water and municipal water application analysis as well as True Turbidity (White Light/InfraRed) direct read method. The SP-910 also contains numerous unique Pyxis Lab direct read and wet chemistry methods not available in other market devices.
Features
Colorimetric & Unique Pyxis Lab® Methods
HACH®, Pyxis Lab® and other solid/liquid reagents are directly compatible for use with the SP-910. Both 10mL and 25mL sample vials are available for use with the SP-910 and all tests offer live graphical display of real-time reading as test occurs. Pyxis Lab® has also developed and provided numerous unique colorimetric and fluorescent test methods integrated into the SP-910 that are not available in other market colorimeters. These include:
Specifications
Item | SP-910 |
Part Number | 50603 |
PTSA | 0-300ppb |
Fluorescein | 0.0-600.0ppb |
Colorimetric Wavelength | 365/420/455/525/560/570/630 nm |
Turbidity Excitation Wavelength | White & IR LED |
Fluorescence Excitation Wavelength | 365/470 nm LED |
Fluorescence Emission Wavelength | 410/525 nm LED |
Wavelength Accuracy | ±1nm |
Absorbance Reproducibility | 0.005au (0-1.5au) (3Sigma) |
Absorbance Linearity Range | 0-1.0au |
Fluorescence Reproducibility | 0.3ppb PTSA / 0.03ppb Fluorescein (3Sigma) |
Fluorescence Detection Limit | 1ppb PTSA / 0.1ppb Fluorescein |
Turbidity Range | 0-200NTU |
Turbidity Detection Limit | 1NTU |
Battery | (4) AA Alkaline Batteries, 6 Months Life |
Display | Graphical LCD 160x240px, Visible under direct sunlight |
Dimensions & Weight | 180x80x38.5mm (7.1×3.1×1.4in) 380g (0.85lbs) without batteries |
Storage Temperature | 0-140 °F (-18-60 °C) |
Operation Temperature | 40-106 °F (4-41 °F) |
Humidity | 85% at 106 °F (41 °C) |
Environmental | IP67 |
Regulation | CE / RoHS |