Pyxis
Pyxis
Pyxis ST-604 Inline Temperature + %Chlorine for Bleach Sensor (0-2%)
The Pyxis ST-600 series inline probes measure the real-time mass/mass concentration of Chlorine in Sodium Hypochlorite Bleach and Chlorine Dioxide (ClO2) in Solution. Unlike typical amperometric sensors for disinfection liquids in ppm level, the ST-600 series inline probes directly measure the optical density of the Bleach or Chlorine Dioxide solution in percentage level without dilution using a UV light source. The ST-600 series probes have a built-in reference light source and a reference light detector in addition to the main light source and main light detector. This configuration ensures a long-term stable measurement of the disinfectant concentration. These probes featuring an expanded detection range and an RTD temperature compensated output signal, allowing the user to measure higher concentrations of industrial Bleach / ClO2 respectively and to obtain a temperature measurement, useful when required to consider the impact of atmospheric temperature conditions on variable concentration readings.
Typical Applications
-
- Biocontrol for industrial water
- Industrial Oxidizer Generation (ie. Hypochlorite /Monochloramine / ClO2)
- Swimming pool and SPA
- Hospital
- Pulp & Paper
Features
- Break-through technology measures disinfection liquids in high concentration
- Direct read technology without dilution or reagent.
- Long maintenance and calibration cycle – 6 months typical.
- Convenient ¾” NPT Inline Tee installation (ST-001) for Large Pump Suction Side
- Convenient ¼” Teflon Tubing Installation for Small Pump Discharge Side
- Probe Cleanliness Diagnostics via uPyxis and Modbus
- Wireless Calibration via uPyxis Mobile or Desktop APP
- Expanded concentration range
- Temperature Compensation (RTD) and Output Signal
Specifications
Item |
ST-604 |
Part Number |
50233 |
Target |
% Chlorine in Bleach |
Concentration Output Range (4-20mA) |
0% - 2% mass/mass |
Temperature Output Range (4-20mA) |
4mA – 0 °C / 32 °F 20mA - 100 °C / 212 °F |
Resolution |
0.01 in percentage |
Accuracy |
±2% of reading or 0.1 percentage, whichever is greater |
Method |
UV Absorbance |
Power Supply |
22-26VDC, 2W |
Outputs |
Isolated 2x 4-20mA Analog Isolated RS-485 Modbus |
Dimension |
Length 6.8-inch, body diameter 1.44 inch |
Weight |
0.37 pounds |
Installation Formats |
¼” OD (7 mm) Teflon tubing for measurement flow ST-001 Inline Tee ¾” CPVC |
Material |
UPVC |
Pressure |
Up to 100 psi (6.9 Bar) |
Temperature |
Operating: 4 °C – 50°C (40-120 °F) Storage: -7 °C - 60 °C (20 - 140 °F) |
Cable Length |
5 feet (Bulkhead Cable w/ IP67 Adapter – 7 pin) 1.5 foot (Flying Lead Cable w/IP67 Adapter – 7 pins |
Calibration |
Two-point calibration against standard solutions |
Rating |
IP67 |
Regulation |
CE / RoHS |