Pyxis
Checkout securely with:
The ST-774 PortaPanel is a prefabricated panel ideally suited for those desiring to use the ST-774 Ultra-Low Dissolved Oxygen Sensor with local display and data acquisition in a portable fashion for oxygen measurement (ie. de-aerator performance studies). The PortaPanel is pre-mounted on an easy-carry, self-standing 316L stainless-steel panel. It arrives with the ST-774 sensor, flow-cell and appropriate water sample/calibration gas flow hardware.
The panel also comes equipped with an integrated UC-50 Display/Data Logging Terminal for connected to the ST-774 via RS-485 Modbus offering live data display, calibration interface and data logging of the sensor. The UC-50 offers a contact relay output and has RS-485 Modbus and (1) 4-20mA outputs for passing the sensor value onto another device. The UC-50 also offers Bluetooth 5.0 capability for those desiring to use the uPyxis mobile or desktop apps for wireless calibration and diagnostics of the Pyxis sensor connected.
Features
Specifications
| Item | ST-774 DO Analyzer |
| Part Number | 53715 |
| Output Range for DO | 0.0-2,000µg/L (ppb) |
| Lower Level of Detection | o.1µg/L (ppb) |
| Resolution | 0.1µg/L (ppb) |
| Accuracy | ±0.3µg/L (ppb) or ±1%, Whichever is Greater |
| Light Source | Blue Light Irradiated Excitation Red Light Reference |
| Response Time (90%) | Liquid Phase <30s / Gas Phase <10s |
| Operational Temperature | 0-50 °C (32-122 °F) |
| Operating Voltage | 22-26VDC, Power 0.6W |
| Signal Output | 4-20mA Analog RS-485 Modbus Digital |
| Dimension | 300mm (11.8in) L x 60mm (2.36in) D |
| Weight | 2150g (4.74lbs) Cable Excluded |
| Installation | 1/4inch OD Swagelok |
| Suggested Flow Range | 50-500mL/min |
| Sample Connection Format | Sample Line & Inlet Stainless Steel / PVC & Polycarbonate |
| Material of Construction | 316L Stainless Steel |
| Working Pressure | 145psi (10Bar) |
| Wet Material | 316L Stainless Steel / PVC & Polycarbonate |
| Cables Provided | MA-1.5CR – 8Pin Female Adapter/Flying Lead – 1.5m Bulkhead Cable with 8Pin Male Adapter – 1.5m |
| Calibration | High Point: o.1% Oxygen in Nitrogen Gas Zero: 99.999% Nitrogen Gas or 5% Catalyzed |
| DO Membrane Cap Life | 2 Years |
| Protection Grade | IP67/IP68 |
| Regulation | CE / RoHS |
Downloads
ST-774 + UC-50 Porta-Panel System Operation Manual