Myron L

Myron L CSAR Low cost Cond/TDS 75 psi @ 60°C with 1000 Ω RTD

No reviews yet
SKU: CSAR
Weight: 1.00 LBS

Call For Quote 1 (916) 315-2691

  • product-security-image-1.jpg
  • product-security-image-2.jpg
  • product-security-image-3.jpg
  • product-security-image-4.jpg

Checkout securely with:

DESCRIPTION

Wetted materials of construction are ABS plastic, with 316 Stainless Steel electrodes. An integral temperature element provides for automatic temperature compensation on both the CSATC and CSAR. The thermowell containing the temperature sensing element in its tip is exposed directly to the fluid stream, assuring rapid response for maximum accuracy.

Potted construction ensures sensor reliability and long service life.

The sensor constant is 1.0. The ranges that can be achieved with these sensors are dependent on the model and range of instrument — up to 20,000 µS/ppm is the normal maximum range of use.

 

FEATURES

  • 3/4” MNPT Fitting allows for easy installation
  • Potted construction - No seals to leak
  • Simple, reliable design for long life
  • Available with or without automatic temperature compensation
  • FDA approved materials

 

SPECIFICATIONS

Maximum Pressure/ Temperature Ratings: 75 psi/5,2 bar at 60°C/140°F

Wetted Materials: FDA Approved

Insulator              ABS

Electrodes           316 Stainless Steel

Temperature Compensation 

CSA: None 

CSATC: 10 KΩ @ 25ºC/77°F

CSAR: 1000 Ω RTD

Connections:     3/4” NPT compatible with BSP 

Process              3/4” MNPT

Electrical            Plain tinned wire ends, 22-24 gauge

Sensor Constant:

Model                    Range                                  Constant

All              0-10 to 0-20,000 µS/ppm                     1.0*